Digital Correlator and
Photon Counter
For Dynamic and Static Light Scattering
|

|
Features at a glance
- USB communications
- Compact size, low power requirements,
less than 300 mW
- Auto- & cross-correlation modes
- Up to 522 hardware channels
- Sampling time 25 ns to 40 ms
- Delay range 25 ns to 1,310 s
- 100% efficient, real-time operation
over the full delay time range
- Supplied with control software for use
with Windows
|
The new TurboCorr digital correlator is the
successor to the BI-9010AT. The new correlator card is designed using a cutting
edge digital signal processor (DSP) which allows all the circuits to be
embedded in a single, high density chip. The benefits of this design include
full, in-system reprogrammability, significantly reduced power consumption as
compared to the BI-9000AT, and a compact design. The TurboCorr now uses USB
communications for enhanced portability and compatability.
The TurboCorr is an advanced and extremely versatile
digital photon correlator and counter designed for use in dynamic (DLS) and
static (SLS) light scattering.
The TurboCorr has an adjustable delay range capable of
producing a contiguous correlation function from 25 ns to 1,310 s representing
a time axis that spans more than 10 decades. The middle and low speed circuits
can be operated as independent correlators, simultaneously processing different
input signals. This can be used, for example, to correlate two light scattering
signals from different detector angles simultaneously. An on-board
microprocessor on each correlator card controls the circuits and collects the
correlation function(s). This relieves the CPU of the host computer of many
tasks. Unlike other correlators, the TurboCorr never requires prescaling.
Operating modes
The TurboCorr digital correlator can be used for real-time
operation in both auto- and cross-correlation modes. It is 100% efficient in
either mode over the entire delay time range. Through the software, the user
can select the number of hardware channels, either linearly spaced or
geometrically spaced. In addition, channel spacing can be user defined. The
TurboCorr can be used for a wide variety of dynamic light scattering
applications.
The TurboCorr includes the BI-DLSW control software for
Windows, shown below. In addition, Brookhaven offers a number of additional
software programs suitable for use with the TurboCorr and the BI-200SM laser
light scattering goniometer system. These applications include particle sizing
by dynamic light scattering (cumulants, NNLS, CONTIN, exponential sampling, and
double exponential), characterization of gels and glass transitions
(Williams-Watts), and molecular weight measurements by total intensity methods
(Zimm, Berry, and Debye plots). A user friendly alignment/stability test
program (BI-IST) is available for the BI-200SM precision goniometer.

Typical Applications
- Particle sizing
- Molecular weights
- Gels
- Glass transitions
- Concentrated colloidal suspensions
- Polymer solutions
Specifications
Communications: USB 2.0
Connections: Four
signal-input connectors and two low voltage outputs.
Channel Layout
|
Section
|
Channels
|
Sampling Time Range, Maximum Time Delay
|
Multiplier
Bits
|
|
|
Low
Speed
|
1
to 240
|
50
us to 40 ms, 1,310 s (40 ms)
|
16
x 16
|
|
|
Mid
Speed
|
1
to 256
|
500
ns to 100 us, 1.6 s (100 us)
|
8
x 8
|
|
|
Std
High Speed
|
14 Channels at 100, 200, 300, 400, 500, 600, 700, 800 ns, 1.0,
1.2, 1.4, 1.6, 2.0 and 2.4 us
|
|
Opt
High Speed
|
12
Channels at 25, 50, 75, 125, 150,175, 225, 250, 275, 325, 350 and 375 ns
|
Maximum Number of Channels: 240 + 256 + 14 + 12 = 522
Delay Time Axis: User selectable (except fixed channels) from
linear, constant ratio, or as individual time points recalled from user files
by name.
Dynamic Range: Over 10 decades (25 ns to 1,310 seconds).
Operation Modes: Automatic and Manual. User definable macro
commands to write, save, and recall unlimited number of sequences for single
keystroke implementation.
Maximum Count Rate: 4 x 107 counts per second.
Standard Input Levels: TTL, 50 ohm input impedance, 5 ns minimum pulse
width. Other levels via, optional, external converter.
Dimensions: 27(H) x 89(W) x 70(D) in millimeters.
Weight: 85 g
Power Requirements: 300 mW per card
Environmental Characteristics: Temperature 10.0 C to 75.0 C,
Humidity 0% to 95%, non-condensing.
A policy of continued improvement may lead to
specification changes
Marked
Options and Accessories
BI-25NS: Twelve, 25 ns
channels.
BI-CON: Stepping motor controller, for use with the BI-200SM goniometer and accessories.
BI-ISTW: Alignment/Stability Software.
BI-ZPW: Zimm plot software.
BI-ISDAW: Particle size distribution software
BI-PCSW: Simple Cumulants (DLS) Software.
BI-SLSW: Particle and fluid structure algorithms: Guinier, Kratky,
Fractal, Rg, ln (Iex) vs q.