Differential Refractometer
|

|
Features at a glance
- Fast instrument warm-up, ~ 1 hour!
- Sensitivity, 2.5 x 10-9 ΔRIU
- Autozero
- Reference cell flush
- 9 measurement ranges
- DNDCW Windows software
- Active temperature control to 80 C
- Wavelength selection: 620, 535 or 470nm
|
Product Description
The BI-DNDC is a deflection type refractometer that may be
purchased for use in either batch or GPC/SEC mode. In batch mode, the specific
refractive index increment, dn/dc, is determined. This value is required as a
parameter in molecular weight measurements using light scattering. In GPC/SEC
mode, dn/dc is already known, and the instrument is used as a concentration
detector for GPC/SEC applications.
Features
All major features are user programmable in order to
achieve the maximum flexibility. The BI-DNDC can be set using the front panel
controls and read using the LED readout or by an external computer. It includes
a microprocessor, which allows user-friendly handling of all functions of the
instrument, as well as refractive index and temperature data transfer to a
personal computer via a standard serial interface.
Most importantly, warm-up time for the BI-DNDC is only one
hour, much less than other models on the market.
Determination of dn/dc
The dn/dc determination requires calibration of the batch
mode instrument with a solvent/solute combination with known dn/dc, typically
KCl in water. The calculated constant is not solvent dependent and may be used
for all further measurements. In batch mode, the instrument comes with BI-DNDCW
software for use with an external PC. Included with the BI-DNDC instrument is a
syringe kit with PEEK and Teflon tubing, fittings and syringes.
Sample may be recovered and reconstituted, if required.
Typically, five or more concentrations, ranging from 0.1 to 10 mg/mL, are
recommended. New instrument readings can be performed in a matter of minutes
since it only takes ~ 0.5 mL of liquid to flush out the previous sample.


Determination of concentration
In this mode, the BI-DNDC serves as a highly sensitive
concentration detector. Even better, with the right wavelength and the BI-MwA,
the BI-DNDC becomes part of a system that eliminates GPC/SEC column
calibration.
Specifications
Cell Angle: 45
Cell Volume: 8 L
Dead Volume (Inlet to Cell/Cell to Outlet): GPC mode sample
cell: 0.25 L, GPC mode reference cell: 100 L; Batch mode sample cell: 100 L;
Batch mode reference cell: 100 L.
Measurement Ranges (ΔRIU Full Scale)(GPC mode): 6x10
-4, 3x10 -4, 1.5x10 -4, 7.5x10 -5,
3.75x10 -5, 1.87x10 -5, 9.3x10 -6, 4.6x10
-6, 2.3x10 -6
n0 range: 1.0 to 1.75 RIU
Sensitivity (GPC mode): 1.5 ng (Fructose in H2O)
Noise Level: < 2.5 x 10 -9, ΔRIU, 25 C
Temperature Accuracy: 0.5 C
Temperature Stability: 0.01 C
Temperature Range: Ambient to 80 C
Specify Wavelength: 620 nm, 535 nm or 470 nm, factory set
Tubing I.D.: GPC mode sample cell: 0.25 mm, GPC mode reference
cell: 0.5 mm; Batch mode sample cell: 0.5; Batch mode reference cell: 0.5 mm.
Flow Rate (typ./max.): 1.0/3.0 mL/min (Analytical Mode),
5.0/12.0 mL/min (Prep. Mode)
Time Constant: 0.1 to 2 seconds
Analog Signal Output: 1 and 10 volt bipolar, full scale
Digital Signal Output: RS-232/COM port
Software: BI-DNDCW for batch mode measurements
Data Transfer Rate (baud): 4800, 9600, 19200, 38400, 57600,
115200
Digital Inputs: Flush, Zero, Sync, Error
Weight: 8 kg
Size: 160 x 175 x 340, mm
A policy of continued improvement may lead to
specification changes.
Options and Accessories
BI-SFTK: Syringes,
fittings and tubing kit