Research Goniometer and
Laser Light Scattering System
|

|
Features at a glance
- Stepping motor controlled, multiangle measuements
- Particle size and size distributions
- Mw, Rg, A2
for dilute polymer solutions
- Cross-correlation detector option
- Avalanche photodiode detector option
- CONTIN, Zimm, Guinier, Fractal, and
more
|
The Brookhaven Approach to Light Scattering
Light Scattering occurs when polarizable particles in a
sample are bathed in the oscillating electric field of a beam of light. The varying
field induces oscillating dipoles in the particles and these radiate light in
all directions. This important and universal phenomena is the basis for
explaining why the sky is blue, why fog and emulsions are opaque and other
observations. It has been utilized in many areas of science to determine
particle size, molecular weight, shape, diffusion coefficients etc. Brookhaven
Instrumentsnull scientists have extensive experience in the development of
instruments and methods for light scattering and this experience has been
incorporated into the BI-200SM to make it the finest instrument available for
research applications of light scattering.
With a Brookhaven Instruments BI-200SM system this rich
field of exploration is open to you for studies of both Static Light Scattering
(SLS) and Dynamic Light Scattering (DLS).
The Bl-200SM Research Goniometer System provides access to
all of these studies with an automatic, modular and versatile system.
It is a precision instrument designed for exacting scattering measurements.
Based on a special turntable with precision ball bearings and stepping motor,
the Bl-200SMnulls modern design and quality construction guarantee precise
measurements due to the wobble-free movement of the detector. It is field
proven in hundreds of laboratories. It is ideal for macro-molecular studies and
submicron particle sizing.
NEW! By taking
advantage of the multiple inputs available on the TurboCorr correlator a new capability of the 200SM system is
created. By adding an optical signal splitter to the beam of scattered light
and a second detector it is possible to eliminate the effects of very high
speed interference from the detector(s). This Cross Correlation System
provides remarkably clean information as seen in the article we have presented.
The BI-200SMnulls modularity simplifies system expansion.
Featuring a standard optical rail, the BI-200SM can be used with neutral
density filters, a polarizer, an analyzer, and a reference detector. Special
sample requirements can often be met by changing the cell holder design or
size.
Special Features of the BI-200SM Research Goniometer
System
Both DLS and SLS modes: Correlation & Intensity Measurements
Large angular Range: 8 to 155 with 25mm cells: 15 to 155 with
12mm cells.
Fine Adjustment Control: Read angle to 0.01 directly on large,
fine-control adjustment knob.
Open Design/Standard Optical Rails: Facilitates user
customizations for special purposes.
Temperature Control: Heating and cooling via any standard
external circulator Separate temperature and filtration plumbing.
Special Design Glass Vats: Specially polished and flat
entrance window ensures minimum flare.
Alignment Cell: Fine-screw vertical adjustment makes finding
the center of rotation easier.
Multiple Laser Line Filters: Filter wheel with 632.8 (HeNe),
514.5/488.0 nm (Ar+), one open position for weak scatterers, and 2 blank
(shutter) positions. Also available with 532 nm bandpass filter
Enhanced Viewing Optics: Coated, precision achromats coupled
with behind-the-slit viewing and high-quality eyepiece make alignment easier.
Additional Alignment Aperture: Alignment of laser much simpler
and faster.
Variety of Cell Holders: High precision cell holders standard,
low-priced round cells available: 27.5 mm dilution vials fit cell holder sleeve. Special small-volume cell available.
Typical Applications
In the SLS mode, time-averaged intensity
measurements are made - at either fixed or variable angles - in the range from
8 to 155 and analyzed with software provided for the methods of Zimm, Berry,
Debye, Guinier, Kratky etc. Such evaluations using measured angular or
concentration dependencies of the intensity of the scattered light provide key
information for those interested in the such topics as:
- Mw Molecular weight determinations
- Rg Radius of gyration
- A2 Determination of second
virial coefficient
- Micro-emulsion technology
- Colloid behavior
- Complex fluid characterization
- Emulsion polymerization
- Particle size growth
- Nucleation processes
The field of DLS measurements is at least
as rich as that of SLS . In this method the dynamics of the
scattered light are determined and analyzed. The short-term intensity
fluctuations (dynamics) of the scattered light arise from the fact that the
scattering particles are undergoing rapid thermal motions. These movements are
called Brownian motion and they cause short term fluctuations in the intensity
of the scattered light. Various terms have been used for this phenomenon. These
are Dynamic Light Scattering (DLS), Photon Correlation
Spectroscopy (PCS) and Quasi-elastic Light Scattering (QELS),
We will adhere to DLS . To uncover the key parameters which
describe the diffusive motions a digital autocorrelator is used to determine
the autocorrelation function (ACF).
From these DLS measurement many interesting subjects may be
explored, among them:
- Particle size distributions
- Particle aggregation phenomena
- Micellar systems
- Micro-emulsion technology
- Colloid behavior
- Vesicles & lipisomes
- Plasmid DNAnulls
- Particle size growth
- Nucleation processes & protein crystallization
Specifications
Beam Focus and Steering: Includes mounts, adjustments, and apertures.
Turntable: 200 mm O.D. turntable, worm gear and ball bearings;
manual or automatic selection of angles in 0.01 steps; precision machined base
with mounting holes; positioning table with micrometer adjustments. (Power
supply and controller for stepping motor is optional)
Specimen Cell Assembly: Including thermal enclosure, centering
adjustments, separate plumbing for temperature control and index matching
liquids, round cell holders, square cell holder and alignment cell.
Index Matching Vat: With optically flat and specially polished
entrance window; precision machined and annealed to minimize stray light.
Periscopic Beam Stop: Located inside vat to prevent flare from
exit window; provided with coupling for use with optional, zero-angle reference
detector.
Main Detector Optics:
- Entrance aperture of 3 mm.
- Coated, adjustable achromat for focusing
scattered light onto 200 micron slit. Slit position adjustable.
- Reflex mirror and high quality eyepiece
for viewing scattering region through slit.
- Filter turret with 632.8, 514.5, 488.0
nm, 2 closed positions, and 1 open position. other filters available.
- Secondary pinhole turret: 100, 200 and
400 micron pinholes for DLS measurements; 1, 2 and 3 mm pinholes for SLS measurements.
A policy of continual improvement may lead to
specification changes.
Options and Accessories
BI-DNDC: Differential
Refractometer. New
BI-CrossCorr: Eliminate detector artifacts by cross
correlation.
BI-SFS: The new BI-SFS sample filtration
system provides a fast and convenient way to clean your light scattering
samples, a necessity for samples smaller than 20nm. Utilizing Teflon tubing and
nonreactive components, the BI-SFS is compatible with most sample filtration
needs. Its modular design allows users to change pump heads and speeds to
encompass virtually any possible application.
BI-APD: Avalanche photo diode, up to 10
times the sensitivity.
BI-HV: High voltage power supply.
BI-LRM: Laser rail and mounts for most lasers. Lasers are
frequently supplied by the user. Contact the factory for recommendations.
BI-TCA: Temperature controller, external circulator type, -20
to 100 C, analog setting and readout, 0.2 stability.
BI-TCD: Temperature controller, external circulator type, -20
to 200 C, alphanumeric setting and readout, 0.1 stability.
BI-FC: Filtration/circulation system for index matching
liquid. New gear pump and rigid Teflon tubing.
BI-RC12: Nominal 12 mm O.D. Round Glass Cell, polyethylene
cap.
BI-SC: 10mm path length Glass Square Cell & PTFE cap.
BI-RC25: Nominal 25 mm O.D. Round Quartz Cell, dual Teflon
tops, mounting ring.
BI-RC27: Nominal 27.5mm O.D. Dilution Vials, package of 100
with HDPE screw caps.
BI-SVC: Small-volume cell for precious samples
BI-2DS: Reference detector including mount. Is recommended for
use with Bl-ZP software when laser drift is greater than 1% and/or when
measuring samples that absorb strongly
BI-PA: Polarization Analyzer, Glan-Thompson prism, 5x10 -6
extinction, 2-position mount.
BI-Spec/Adap Inquire about adapting parts of your existing
light scattering system.
|

|
|
Legend
1. Laser rail and mounts, optional
2. Precision-machined base
3. Turntable
4. Rigid rotating arm
5. Upright for 2 mm alignment aperture
6. Detector rail
7A,B. Center of rotation adjustment screws
7C. Center of rotation locking screw
8. Center of rotation adjustment table
9. Beam focusing and steering lens assembly
10. Sample cell assembly
11. Beam Stop
12A. Lens adjustment, horizontal
12B. Lens adjustment, vertical
13A. Slit adjustment, horizontal
13B. Slit adjustment, vertical
14. Mirror adjustment
15. Eyepiece
16. Pinhole wheel
17. Filter wheel
18. Photomultiplier housing
19. Angle adjustment
20. Clutch release
21. Support ring
|